Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 40

Advertising
background image

4 Operation

Model 3010TB

4-12: Part I

Teledyne Analytical Instruments

If you are using password protection, you will need to enter your

password to gain access to either of these functions. Follow the instructions
in sections 4.3.3.2 or 4.3.3.3 to enter your password. Once you have gained
clearance to proceed, you can enter the

Zero

or

Span

function.

4.4.1 Zero Cal

The

Zero

button on the front panel is used to enter the zero calibration

function. Zero calibration can be performed in either the automatic or manual
mode. In the automatic mode, an internal algorithm compares consecutive
readings from the sensor to determine when the output is within the accept-
able range for zero. In the manual mode, the operator determines when the
reading is within the acceptable range for zero. Make sure the zero gas is
connected to the instrument. If you get a CELL FAILURE message skip to
section 4.4.1.3.

4.4.1.1

Auto Mode Zeroing

Press

Zero

to enter the zero function mode. The screen allows you to

select whether the zero calibration is to be performed automatically or manu-
ally. Use the

Δ∇ arrow keys to toggle between AUTO and MAN zero

settling. Stop when AUTO appears, blinking, on the display.

Zero: Settling: AUTO
<ENT> To Begin

Press

Enter

to begin zeroing.

#### PPM Zero
Slope=#### ppm/s

The beginning zero level is shown in the upper left corner of the dis-

play. As the zero reading settles, the screen displays and updates information
on Slope (unless the Slope starts within the acceptable zero range and does
not need to settle further).

Then, and whenever Slope is less than 0.08 for at least 3 minutes,

instead of Slope you will see a countdown: 5 Left, 4 Left, and so fourth.
These are five steps in the zeroing process that the system must complete,
AFTER settling, before it can go back to

Analyze

.

#### PPM Zero
4 Left=### ppm/s

The zeroing process will automatically conclude when the output is

within the acceptable range for a good zero. Then the analyzer automatically
returns to the

Analyze

mode.

Advertising