Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 97

Advertising
background image

A-9

Oxygen Analyzer

Appendix

Teledyne Analytical Instruments

BY-PASS:

To improve the system response, a by-pass can be added to increase

the sample flow rate to the analyzer by a factor of ten. A by-pass provides a
sample flow path around the analyzer of 2 - 18 SCFH. typically.

CALIBRATION GAS:

3000 series analyzer requirements for units with Auto-Cal options. The

customer must supply a control valves (or restrictors) for any SPAN or ZERO
gas source which is attached to the Auto-Cal ports. The valve should be ad-
justed to the same flow rate as the sample gas . When restrictors are used, the
gas pressure must be adjusted to achieve the proper flow rate.

OPERATION WITHOUT A RESTRICTOR DEVICE:

Operation without a restrictor device is not recommend as mentioned

above. A 3010TB without any flow restrictor device was tested on 11-19-97.
This results in a flow rate of 2.4 SLPM @ 1 PSIG. This is a cv of 0.023 for the
standard sample sys.

REFERENCE: FLOW_1.XLS & FLOW_2.XLS for information on

flow rates at various pressures.

TAI PART NUMBERS

RESTRICTOR KIT:

A68729

UNION (SS)

U11

LP. RESTRICTOR

R2323 ( LOW PRESSURE / VAC.
SERVICE )

STD.. RESTRICTOR

R2324

BLUE DOT

NUT

N73

FERRULE

F73

FERRULE

F74

BOTH FERRULES
ARE REQUIRED

CONVERSIONS:

1 PSI

=

2.04 INCHES OF MERCURY (in. hg.)

1 SCFH =

0.476 SLPM

Advertising