Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 67

Advertising
background image

Part II: 2-5

Oxygen Analyzers

Part II: Analysis Units

Teledyne Analytical Instruments

Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel Cell

2.2.6 Micro-Fuel Cell “Class”

Analytical Instruments manufactures Micro-Fuel Cells with a variety of

characteristics to give the best possible performance for any given applica-
tion. Each unique cell type is given an alphanumeric designation consistent
with its application. This designation is its “class.”

2.2.6.1 Hydrogen and/or Helium Service

If the sample gas contains 10 % or more hydrogen and/or helium, only

“clamp” cells are used. These Micro-Fuel cells are identified by the suffix -C
added to the cell class number.

2.2.6.2

Class L-2C Cell

The class L-2C cell is for general purpose, and hydrogen/helium ser-

vice, trace oxygen analysis, where the slightly larger size of the cell is ac-
ceptable to obtain a longer life than is available with the standard size cell.

Nominal output in air is 0.5 mA, and 90 % response time is 7 s.

Expected life is 12 months.

Advertising