Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 26

Advertising
background image

3 Installation

Model 3010TB

3-10: Part I

Teledyne Analytical Instruments

The concentration in percent

The range is use (HI< MED< LO)

The span of the range 0-100%, etc)

Which alarm - if any - are disabled (AL-x DISABLED)

Which alarms - if any - are tripped (AL-x ON)

Each status output is followed by a carriage return and line feed.

Three input functions using RS-232 have been implemented to date.

They are described in Table 3-4.

Table 3-4: Commands via RS-232 Input

Command

Description

as<enter>

Immediately starts an autospan.

az<enter>

Immediately starts an autozero.

st<enter>

Toggling input. Stops/Starts any status message output
from the RS-232, Until st<enter> is sent again.

The RS-232 protocol allows some flexibility in its implementation.

Table 3-5 lists certain RS-232 values that are required by the 3010TB.

Table 3-5: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds

Remote Sensor and Solenoid Valves: The 3010TB is a single-chassis

instrument. However, the REMOTE SENSOR and SOLENOID RETURN
connectors are provided for use with a remote sensor and/or sampling sys-
tem, if desired. See Figure 3-7 and 3-8.

Advertising