22: part i – Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 50

Advertising
background image

4 Operation

Model 3010TB

4-22: Part I

Teledyne Analytical Instruments

IMPORTANT: In the event of loss of flow through the analyzer, if the

vent is vented to a location of high oxygen content,
oxygen will back diffuse through the vent line and in
most cases quickly saturate the cell with oxygen which
can then require a quite long purge down time for
the sensor when then exposed to low oxygen concen-
trations. In the event that flow is to be interrupted into
the analyzer, it is suggested that the user do one of the
following:

1.

Bag the sensor in nitrogen during this time

2.

Install a shut off valve on the vent port of the ana-
lyzer or somewhere within the users sample system.