Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 34

Advertising
background image

4 Operation

Model 3010TB

4-6: Part I

Teledyne Analytical Instruments

TRAK/HLD Auto—Cal
PSWD Logout More

Use < > arrows to blink

Auto—Cal

, and press

Enter

. A new screen for

Span/Zero set appears.

Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h

Press < > arrows to blink

Span

(or

Zero

), then press

Enter

again. (You

won’t be able to set OFF to ON if a zero interval is entered.) A Span
Every ... (or Zero Every ...) screen appears.

Span Every 0 d
Start 0 h from now

Use

ΔΔΔΔΔ∇ arrows to set an interval value, then use < > arrows to move to

the start-time value. Use

ΔΔΔΔΔ∇ arrows to set a start-time value.

To turn ON the Span and/or Zero cycles (to activate Auto-Cal): Press

System

again, choose

Auto—Cal,

and press

Enter

again. When the Span/

Zero values screen appears, use the < > arrows to blink the Span (or Zero)
OFF/ON field. Use

ΔΔΔΔΔ∇ arrows to set the OFF/ON field to ON. You can

now turn these fields ON because there is a nonzero span interval defined.

4.3.3 Password Protection

If a password is assigned, then setting the following system parameters

can be done only after the password is entered: span and zero settings,
alarm setpoints, analysis range definitions, switching between autoranging
and manual override, setting up an auto-cal, and assigning a new password.
However, the instrument can still be used for analysis or for initiating a self-
test without entering the password.

If you have decided not to employ password security, use the default

password TETAI. This password will be displayed automatically by the
microprocessor. The operator just presses the Enter key to be allowed total
access to the instrument’s features.

NOTE: If you use password security, it is advisable to keep a copy of

the password in a separate, safe location.

4.3.3.1

Entering the Password

To install a new password or change a previously installed password,

you must key in and

ENTER

the old password first. If the default password

Advertising