WRM-10P AND WRM-40 USER’S MANUAL

REV 3

36

3.6

Performing a Special Resistance Test

The Special Resistance Test is used to conduct a resistance test for a pre-defined period ranging
from 1 to 45 minutes. The resistance data is recorded at one minute intervals. Use the steps
below to perform a Special Resistance Test:

a. Start from the “START-UP” menu:

Press the

[1]

key (TEST XFMR).

b. The following screen will be displayed:

Press the

[3]

key (SPECIAL TEST V1, V2) to perform a V1 and V2 special resistance test.

Press the

[4]

key (SPECIAL TEST V1) to perform a V1 special resistance test.

c. The following screen will be displayed:

Enter the test time (between 1 to 45 minutes) using the alpha-numeric keypad, and
then press the

[ENTER]

key.

d. The following screen will be displayed:

1. YES

Press the

[1]

key (YES) if you wish to convert readings to standard temperature.

The following screen will be displayed:

WINDING MATERIAL:
1. COPPER, Tk=234.5

2. ALUMINUM, Tk=225

3. MANUALLY ENTER Tk

CONVERT READINGS

TO STANDARD TEMP?

1. YES

2. NO

ENTER SPECIAL TEST
TIME MINUTES (1-45):

_

1. V1 & V2 TEST

2. V1 ONLY TEST

3. SPECIAL TEST V1,V2

4. SPECIAL TEST V1

1. TEST XFMR 11/02/09

2. SETUP 12:33:01

3. USER DIAG