0 troubleshooting guide – Vanguard WRM-40 User Manual

Page 61

Advertising
background image

REV 3 WRM-10P AND WRM-40 USER’S MANUAL

57

5.0 Troubleshooting

Guide

PROBLEM DESCRIPTION

Probable Cause

Suggested Solution

No current displayed on
the LCD screen when
performing a test.

Opened current cable connection.

Opened transformer windings.

• Check current cable connection.
• Perform a quick test using a shunt

or a piece of aluminum bar.

Resistance reading is zero
or invalid while current
value is displayed on the
LCD screen.

• Opened sense leads connection.
• Sense leads problem.

• Check sense leads connection to

D.U.T.

• Check sense leads using an

ohmmeter.

• Perform a diagnostic test. See

section 3.8.