Vanguard WRM-40 User Manual

Page 41

Advertising
background image

REV 3 WRM-10P AND WRM-40 USER’S MANUAL

37

Press the

[1]

key (COPPER, Tk=234.5) to select copper as the winding material.

Continue to step d.

Press the

[2]

key (ALUMINUM, Tk=225) to select aluminum as the winding

material. Continue to step d.

Press the

[3]

key (MANUALLY ENTER Tk) to manually enter the Tk value. The

following screen will be displayed:

You can increase the Tk value by 0.5

°C increments by pressing the

[PAPER

Contrast]

key.

You can decrease the Tk value by 0.5

°C increments by pressing the

[PAPER

Contrast]

key.

Press the

[ENTER]

key to confirm the temperature setting. Continue to step e.

2. NO

Press the

[2]

key (NO) if you do NOT wish to convert readings to standard

temperature. Continue to step e.

e. The following screen will be displayed:

Use the

[PAPER Contrast]

and

[PAPER Contrast]

keys to adjust the D.U.T.

(Device Under Test) temperature.

Press the

[ENTER]

key to confirm the D.U.T. temperature setting.

f. The following screen will be displayed:

Use the

[PAPER Contrast]

and

[PAPER Contrast]

keys to adjust the

reference temperature.

Press the

[ENTER]

key to confirm the reference temperature setting.

REFERENCE TEMP:

85.0

°C 185.0°F

↑↓ to adjust temp

“ENTER” to accept

D.U.T. TEMP

25.0

°C 77.0°F

↑↓ to adjust temp

“ENTER” to accept

Tk:

230.0

°C

↑↓ to adjust Tk

“ENTER” to accept