Vanguard WRM-40 User Manual

Page 8

Advertising
background image

WRM-10P AND WRM-40 USER’S MANUAL

REV 3

4

1.2 Technical

Specifications

1.2.1. WRM-10P Technical Specifications

Table 1. WRM-10P Technical Specifications

TYPE

Portable transformer winding resistance meter

PHYSICAL SPECIFICATIONS

16.8”W x 12.6”H x 10.6”D (42.6 cm x 32.0 cm x 26.9 cm); Weight: 27 lbs
(12.2 kg)

INPUT POWER

100

120 Vac or 200

240 Vac (factory pre-set), 50/60 Hz

(Please see the voltage setting notes in section 2.1)

RESISTANCE READING

RANGE

1 micro-ohm

2,000 ohms

ACCURACY

1 – 19,999 micro-ohms: ±0.5% reading, ±1 count;
20 – 999 milliohms: ±1% reading, ±1 count;
1 – 2,000 ohms: ±1.5% reading, ±1 count

TEST VOLTAGE

36 Vdc max

TEST CURRENT RANGE

Auto range, 10 Amperes max

DISPLAY

Back-lit LCD Screen (20 characters by 4 lines); viewable in bright sunlight
and low-light levels

PRINTER

2.5-inch wide built-in thermal printer

KEYPAD

Rugged membrane keypad (10 alpha-numeric keys, 6 function keys)

INTERNAL TEST RECORD

STORAGE

Stores 63 test records of 48 readings each

COMPUTER INTERFACE

One RS-232C (19,200 Baud) port

PC SOFTWARE

Windows

®

XP/Vista-based software is included with purchase price

SAFETY

Designed to meet IEC61010 (1995), UL61010A-1, CSA-C22.2 standards

ENVIRONMENT

Operating: -10˚C to 50˚ C (15˚F to +122˚ F); Storage: -30˚ C to 70˚ C (-
22˚F to +158˚ F)

CABLES Three 50-foot test cables, ground cable, power cord and cable bag

OPTIONS Transportation case

WARRANTY One year on parts and labor

NOTE

The above specifications are valid at nominal operating voltage and at a
temperature of 25

°C (77°F). Specifications may change without prior notice.

Advertising