4.INSTALLATION

IM 1E8B0-01E

4-5

IMPORTANT

(13)Grounding

Improper grounding can have an adverse affect on the flow measurement.
Please ensure that the instrument is properly grounded.

The electromotive force of the magnetic flowmeter is minute and it is easy to be affected
by noise. And also that reference electric potential is the same as the measuring fluid
potential. Therefore, the reference electric potential (terminal potential) of the Flow Tube
and the Converter/Amplifier also need to be the same as the measuring fluid. And
moreover, that the potential must be the same with ground.

Please be sure to grounding according to Figure 4.1.7.

Earth ring

600 V vinyl insulated
electric cable
(2mm

2

or larger)

Grounding resistance: 100

or less

(10

or less for TIIS(JIS) Flameproof type)

Note: See “4.4.1 Protective Grounding” for

information on protective grounding.

Figure 4.1.7

Gounding