Yokogawa ADMAG CA User Manual

Page 35

Advertising
background image

IM 1E8B0-01E

5-6

5. BASIC OPERATING PROCEDURES

Item

Display

Content

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

Press the SET key twice to enter data.
(All data will start flashing when the key is pressed the first time.
Then wait 2 to 3 seconds before pressing the key the second time.)

Hold down the SHIFT key and press the SET key to switch to
the flow rate data display.

Data Input

Switching to Flow
Data Entry

First time

Second time

%

(3) Procedures for Changing Decimal Place

Before starting this procedure check in the data list if the position of the decimal point

for the desired parameter decimal places can be changed.

Item

Display

Content

SET SHIFT INC

Press the SHIFT key to move the flashing segment to the decimal
point.

Selecting
Decimal
Point

SET SHIFT INC

Press the INC key to move the decimal point to the right.

Moving
Decimal
Point

SET SHIFT INC

Press the SET key twice to enter data.
(All data will start flashing when the key is pressed the first time.
Then wait 2 to 3 seconds before pressing the key the second tme.)

Data Entry

SET SHIFT INC

First time

Second time

Example of parameter change : Changing the span from 1.000 to 10.00