4.INSTALLATION

IM 1E8B0-01E

4-7

4.2.2 Flowmeter piping

Mis-aligned or slanted piping can lead to leakage and damage to flanges.

Please be sure the inner diameter of the gasket between ADMAG CA and
piping flange does not protrude to inner piping. It can lead to error in
measurement. This is important especially for low conductivity fluid.

• Please correct mis-alignment or slanted piping and improper distance between

mounting flanges before install the flowmeter. (Please refer to Figure 4.2.2)

• Inside a pipeline which is newly installed, some foreign substances (such as welding

scrap or wood chips) may exist. Please remove them by flushing piping before
mounting flow tube.

Mis-alignment

slant

Figure 4.2.2

Slant and Mis-alignment of Flowmeter Piping

4.2.3 Alteration of LCD Display Orientation

LCD display orientation can be altered according to piping configurations if horizontal
or vertical, just by removing four screws, adjusting unit orientation and fixing the screws
tightly again as shown in Figure 4.2.3.

Figure 4.2.3

Procedure of Altering LCD Display Orientation

Orientation of display unit is limited to the two positions shown in this figure.

CAUTION

IMPORTANT

NOTE