Yokogawa ADMAG CA User Manual

Page 50

Advertising
background image

IM 1E8B0-01E

6-13

6. FUNCTION AND DATA SETTINGS

• Signal processing procedures:

The function sets a certain upper and lower limit (rate limit value) for first order delay

response values of flow rate data obtained in a previous sampling. If currently
sampled flow rate data exceeds or goes below the limit is regarded as current flow
rate value. Signals whose protruding portions show the same trends during a certain
number of sampling times (dead time) are identified as flow rate signals.

Example 1: Step Input

1%

10%

Step Signal

Flow rate value
after rate limiting

Flow rate value after
damping

(3)

(1)

(2)

9

18

(4)

63.2%

Input: 0 to 10%
Damping time constant: 3 sec.
Dead time: 3 sec.
Rate limit value: 1%

Dead time: 3 sec.

Sampling times

(1) Shows 1% response cause of excessive signal beyond the rate limit. However, actual

output is under damping that described by a solid line.

(2) Shows the flow rate signal (1%) of just after damping calculation (1) and rate limit

value.

(3) This signal is recognized as a flow rate signal since it does not return to within the

rate limit value within the dead time.

(4) The output signal follows the damping curve and tracks the step signal.

Example 2: Slurry Noise

Slurry Noise

Flow rate value after rate limiting and damping.

Flow rate value after
rate limiting

Dead time: 1 sec.

Time

+1%

–1%

Input: 0 to 10%
Damping time constant: 1 sec.
Dead time : 1 sec.
Rate limit value: 1%

The left figure shows that the slurry
noise is not a flow rate signal.

Display

Description

Set a rate limit function value in parameter number "N3".

Range of possible settings : 0 to 10%.

set the dead time in parameter number "N4".

Range of possible settings : 0 to 15 seconds.

Default: 5%

Default: 0%

(rate limit function off)

*Item "n" can be opened by entering "55" in parameter number "L2".

Advertising