10.1 Standard Specifications

Degrees of Protection

: IP67, NEMA 4X, JIS C0920 Water tight

protection

Grounding

: 100

or less

* In case of TIIS(JIS) Flameproof type, JIS Class C

(10

or less) or JIS Class A (10

or less)

* In case of explosion proof type except TIIS, follow the domestic electrical
requirements as regulated in each country.

10.1.1 Magnetic Flow Converter

A handheldterminal is necessary to set parameters for models with no
indicator.

Output Signal

:

Current output

; 4 to 20mA DC (Load resistance 0 to 750

)

Pulse output / Alarm output

; Transistor contact (contact rating: 30V DC(OFF),

200mA(ON))

Communication signals (optional) :

BRAIN

communication signals (superimposed on a 4 to 20mA DC signal)

Load resistance

; 250 to 600

(including cable)

Load capacity

; 0.22

µ

F or less (including cable)

Load inductance

; 3.3mH or less (including cable)

Maximum cable length

; 2km (6560 ft) (using CEV cable)

Space from power line

; 15cm or more (parallel wiring should be

avoided)

Input Impedance of Receiver Connected to the Receiving Resistance;

10k

or larger (at 2.4kHz)

Span Setting Functions

: Volumetric flow setting is possible by setting

volume unit, time unit, flow rate value and flow
tube size.

Volume units

; m

3

, L(liter), cm

3

, gallon(us), barrel (=158.987L)

Velocity units

; m, ft

Time units

; s (sec), min, h (hour), d (day)

Flow Tube Size

; mm, inch

Instantaneous Flow Rate Display Function:

Flow rate can be displayed either in engineering
units or percent of span.

Totalizer Display Function

: Totalized volume in engineering unit can be

displayed by setting a totalizing factor.

Pulse Output Function

: Pulse output after scaling can be output by

setting a pulse factor.

Pulse width

; Duty 50% or fixed pulse width (0.5, 1, 20, 33,

50 or 100ms)

Output rate

; 0.0001 to 1000PPS

NOTE