H) • span value setting – Yokogawa ADMAG CA User Manual

Page 40

Advertising
background image

IM 1E8B0-01E

6-3

6. FUNCTION AND DATA SETTINGS

(2) Span Setting using Display Keys (Example setting: 30.00 m

3

/h)

Span Value Setting

Switch Operation

Display

Description

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

Press the SET key during normal flow rate data display
to display the data mode shown in the left figure.

Press the SHIFT key to cause the second digit from
the left to flash.

Press the INC key to call up parameter number 03.

Press the SET key to move the flashing segment to
the first digit in the data area.

Press the INC key to set the first digit to 0.

Press the SHIFT key to move the decimal point.

Press the INC key to move the decimal point to "000.00".
*To set a "00000." as the data area is a four-digit LCD,
the three digits and the space will be displayed as "000".

Press the SHIFT key to move the flashing segment two
digits to the right.

Press the INC key to enter "3".

Press the SET key once to cause all the data to flash.
*As the data display now will be scrolled, it may not
appear as shown in the left figure.

Wait 2 to 3 seconds when the data starts flashing. Then
press the SET key to return the flashing segment to the
leftmost digit.
*This display will also be scrolled to display 5-digit data.

Default is set as 1.0000. Change this into 030.00.

(Setting is now completed)

Note : 5-digit span data can be displayed, but only digits will be supplyed at one time. As a result, when the

last digit is changed, the last digit and the space will be displayed as the last three digits. (See 5.2.2
"Display of 5-digit data".)