HT instruments SIRIUS87 User Manual

Page 15

Advertising
background image

SIRIUS

87

EN - 13

4.1.2.1. Mode

"AUTO"

 At the end of the test, if the average

resistance value Ravg results to be
lower than 5

 the instrument emits a

double sound signal indicating the
positive outcome of the test and
displays one screen similar to the
screen alongside.

CAL

LOW



1.07

219

mA

AUTO


SAVING:

The test can be stored pressing the

SAVE key twice

(according to § 5.1).



4.1.2.2. Mode "R+TIMER" or “R-TIMER"

 If during the test a resistance value

R+Timer or R–Timer lower than 5

was detected, the instrument (after the
set time has elapsed) emits a

double

sound signal indicating the positive
outcome
of the test displays one screen
similar to the screen alongside.

CAL LOW



1.08

219

mA

5

s

R+
R-

TIMER


SAVING:

The test can be stored pressing the

SAVE key twice

(according to § 5.1).

Average test
current value
Iavg.

Average
resistance value
Ravg.

Value of the test
current

I+ or I-.

The symbols

R

+

or

R

-

are displayed.

Duration time of
the test.

SAVE

SAVE