9. TECHNICAL SPECIFICATIONS

9.1. TECHNICAL

FEATURES

Accuracy is indicated as [% reading + (number of digits) * resolution]. It refers to the
atmospheric conditions listed in § 9.2.1.


- Continuity (LOW

)

Test mode

Measuring range (

)

Resolution (

)

Accuracy

AUTO, R+TIMER, R-TIMER

0,01 – 19,99

0,01

(2% rdg + 2 dgt)

20,0 – 99,9

0,1

Test current

>200mA DC up to 5

 (include the resistance of the calibration)

Test current resolution

1mA

Open circuit Test Voltage

6V < V

0

< 12V


- Insulation resistance (R

ISO

)

Test voltage (V)

Measuring range (

)

Resolution (

)

Accuracy

50

0,01 - 19,99

0,01

(2% rdg + 2 dgt)

20,0 - 49,9

0,1

50,0 - 99,9

0,1

(5% rdg + 2 dgt)

100

0,01 - 19,99

0,01

(2% rdg + 2 dgt)

20,0 - 99,9

0,1

100,0 - 199,9

0,1

(5% rdg + 2 dgt)

250

0,01 - 19,99

0,01

(2% rdg + 2 dgt)

20,0 - 199,9

0,1

200 - 249

1

250 - 499

1

(5% rdg + 2 dgt)

500

0,01 - 19,99

0,01

(2% rdg + 2 dgt)

20,0 - 199,9

0,1

200 - 499

1

500 - 999

1

(5% rdg + 2 dgt)

1000

0,01 - 19,99

0,01

(2% rdg + 2 dgt)

20,0 - 199,9

0,1

200 - 999

1

1000 - 1999

1

(5% rdg + 2 dgt)

Automatic selection of the measuring range
Open circuit voltage

Rated voltage test –0% +10%

Short circuit current

<6,0mA with 500V set

Rated measuring current

>2,2mA with 500V on 230k

1mA a 1K

 x V

NOM

(

 500 V)


- RCD tripping tests (RCD

or RCD

)

RCD type

AC, A General and Selective

Range of phase to earth voltage

100

 265V 50Hz + 0.5 Hz

Rated tripping currents (I

N

)

10mA, 30mA, 100mA, 300mA, 500mA


- Contact voltage (U

t

)

Measuring range (V)

Resolution (V)

Accuracy

0

 2U

t lim

0,1

-0%, +(5% rdg + 3 dgt)

U

t

lim (U

L

): 25V o 50V