HT instruments SIRIUS87 User Manual

Page 19

Advertising
background image

SIRIUS

87

EN - 17

4.2.1.1. Mode

"MAN"

6. Press

the

START/STOP key.



The instrument effects the test lasting:

 Minimum 6 seconds in case the key is pressed and released within 5 seconds.
 Until the key is released for all the other cases.

 At the end of the test, in case the

resistance value R

ISO

detected results

to be lower than R

MAX

(depending on

the selected voltage see following table)
and the test is effected at the
selected rated voltage value
, the
instrument emits a

double sound

signal

indicating the positive

outcome of the test displays one
screen similar to the screen alongside.

R

ISO



1.00

M

500 V 20

s


MAN



The values of measured resistance of isolation always must be confronted with the
normative limits (see table) for being able to assert if the system is to norm.

SAVING:

The test can be stored pressing the

SAVE key twice

(according to § 5.1).



Rated voltage value
selected for the test.

Test duration.
In this case the
START/STOP
key has been
pressed for 20
seconds.

Manual mode.

SAVE

R

ISO

insulation

resistance value

.

START

STOP