KROHNE IFC 110 Converter User Manual

Page 18

Advertising
background image

Sect. 2.3.1

Part A System installation and start-up

18

IFC 110 F

05/2003

2.3

Pulse outputs P and A1


2.3.1 Pulse output P for electronic totalizers (EC)

Pulse output P is electrically isolated from all other circuits.


Setting data and functions can note down in the Table in Sect. 3.

Please also refer to Section 2.7 ”Standard factory settings”.


All operating data and functions are adjustable, see Sections 4.4 and 5.7, Fct. 1.05.


Active mode:

uses the internal power supply, terminals E+/E-

Passive mode:

requires external power supply, U

ext

≤ 32V DC/24V AC, I ≤ 30mA


Max. adjustable frequency 10 kHz


Scaling

in pulses per unit time (e.g. 1000 pulses/s at Q

100%

flow) or

in pulses per unit volume (e.g. 100 pulses/m³ or US Gal).


Pulse width

symmetric, pulse duty factor 1:1, independent of output frequency,

automatic, with optimum pulse width,

pulse duty factor approx. 1:1 at Q

100%

, or

pulse width range from 0.01 to 1 s adjustable as required for

correspondingly lower output frequency.


Forward/reverse flow measurement (F/R mode) is possible, see Section 5.15.


Connection diagrams see Section 2.6


Schematic wiring diagram for pulse output P for electronic totalizers EC

Similar to a relay contact, this pulse output switches direct and alternating voltages.