KROHNE IFC 110 Converter User Manual

Page 37

Advertising
background image

Part B IFC 110 F Signal converter Sect. 4.4

05/2003

IFC 110 F

37

Fct. Text

Description

and

setting

3.05

USER UNIT

Set any required unit for flowrate and counting

TEXT VOL.

Set text for required flowrate unit (max. 5 characters)

Factory setting = Liter or US MGal

Characters

which

can be assigned to each place:

● A-Z, a-z, 0-9, or “ – “ (= blank character).

Press

key to transfer to subfunction “FACT. VOL.”

FACT. VOL.

Set conversion factor (F

M

) for volume

Factory setting ”1.00000 E+3” for ”Liter” or ”2.64172E-4” for ”US MGal”

(exponent notation, here 10³ or 2.64172 x 10

-4

)

Factor

F

M

= volume per 1m³.

Setting

range

● 1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

key to transfer to subfunction “TEXT TIME”.

TEXT TIME

Set text for any time (max. 3 characters)

Factory setting = ”h” (hours)

Characters

which

can be assigned to each space:

● A-Z, a-z, 0-9, or “ – “ (= blank character).

Press

key to transfer to subfunction “FACT. TIME”

FACT. TIME

Set conversion factor (F

T

) for time

Factory setting ”3.60000 E+3” for ”h” (exponent notation, here 3.3 x 103).

Set factor F

T

in seconds.

Setting

range

● 1.00000 E-9 to 9.99999 E+9 (= 10

-9

to 10

+9

)

Press

key to return to Fct. 3.05 “USER UNIT”.

3.06

APPLICAT.

Set modulation range of A/D converter

FLOW

● STEADY (150% of Q

100%

)

● PULSATING (1000% of Q

100%

)

Press

key to change to subfunction ”EMPTY PIPE”.

EMPTY PIPE

Empty pipe detection EPD (see Sect. 6.9)

● NO

(Press

key to change to subfunction “ADC Gain“)

● YES

(Press

key to change to “VAL. FULL“)

● VAL. FULL

(Press

key, precautinory query)

● CALIB. NO

(Press

key to change to “VAL. EMPTY“)

● CALIB. YES

(Press

key, calibration will start with flashing display “WAIT“,

duration approx. 20 seconds)

Make sure that measuring tube is completely filled!

● STORE NO

(Press

key to change to “VAL. EMPTY“)

● STORE YES (Presskey to change to “VAL. EMPTY“)

● VAL. EMPTY (Presskey, precautinory query)

● CALIB. NO

(Press

key to change to “VAL. EMPTY“)

● CALIB. YES

(Press

key, calibration will start with flashing display “WAIT“,

duration approx. 20 seconds)

Make sure that measuring tube is completely empty!

● STORE NO

(Press

key to change to subfunction

ADC

GAIN“)

● STORE YES (Presskey to change to subfunction

“ADC

GAIN“)

Please Note:

The values of the measured impedances must be in range 0 - 150.

The difference of the value VAL. EMPTY must be

10 greater than the value of VAL. FULL!

ADC GAIN

Set gain of A/D converter

● AUTO

● 10

● 30

● 100

Select with key

↑ or ↓

Press

key to change to subfunction ”SPEC. FILT.”.

SPEC. FILT.

Activate special filter for noise/interference suppression?

PLEASE NOTE information and examples given in Sect. 6.6.

● NO

(Press

key to change to Fct. 3.06 ”APPLICAT.”)

● YES (Press key to change to subfunction ”LIMIT VAL.”).


Advertising