KROHNE IFC 110 Converter User Manual

Page 60

Advertising
background image

Sect. 6.4

Part C Special applications, functional checks, service, and order numbers

60

IFC 110 F

05/2003


6.4.3 KROHNE RS 485 Interface (Option)

Electrical connection

RS 485 Interface (Option, connection at plug-in terminal RS)

*

The shields can be grounded as follows:

1) directly at both ends
or where there is risk of circulating currents,

2) directly at one end (static shielding) or

3) directly at one end and capacitive at the other end.

In the case of 2) and 3), a decrease in EMC is to be expected.


It is essential to blank off the final signal converter's electrical bus. To do this, solder up the
semicircular circuits of solder points S3 and S4 on the RS485 - PCB. For further information refer
to section 8.7.

The RS 485 protocol is available. Please contact your local KROHNE supplier / company.


Settings for use in interface operation

Fct.

Parameter

KROHNE RS 485 Interface

3.09

Communication KROHNE

Address

000-239

Baud

rate

• 1200

• 9600

• 2400

• 19200

• 4800

For further information on setting the signal converter refer to chapters 4 and 5.