KROHNE IFC 110 Converter User Manual

Page 5

Advertising
background image

05/2003

IFC 110 F

5

System description


Electromagnetic flowmeters are precision instruments designed for linear flow measurement of
liquid products

The process liquids must be electrically conductive:,

≥ 5 µS/cm

(for cold demineralized water

≥ 20 µS/cm).

The full-scale range Q

100%

can be set as a function of the meter size: flow velocity of 0.3 - 12 m/s

or 1 - 40 ft/s (s. Section 10.4.).

Product liability and warranty


The electromagnetic flowmeters are designed solely for measuring the volumetric flowrate of
electrically conductive, liquid process products.

These flowmeters are available for use in hazardous areas.
Special regulations apply in this case, which are given in the special EEx directions.

Responsibility as to suitability and intended use of these electromagnetic flowmeters rests
solelywith the operator.

Improper installation and operation of the flowmeters (systems) may lead to loss of warranty.

In addition, the “General conditions of sale” forming the basis of the purchase contract are
applicable.

If flowmeters need to be returned to KROHNE, please note the information given on the last-but-
one page of these Instructions. KROHNE regrets that it cannot repair or check your flowmeter(s)
unless accompanied by the completed form sheet.

CE / EMC / Standards / Approvals


The here described Electromagnetic flowmeters meet the NAMUR Directive NE21,
the protection requirements of Directive 89/336/EEC in conjunction with EN
61326-1 (1997) and A1 (1998), as well as Directives 73/23/EEC and 93/68/EEC
in conjunction with EN 61010-1, and bear the CE marking.

Software history

Display & control unit

Amplifier (ADC)

Inputs and outputs (I/O)

Software Status Software Status Software Status

3.19937.02.00

current

8.13393.02.00

current

3.16230.01.00

current




IMPORTANT!
In respect of EEx versions, pay regard to all directions marked with the

symbol,

and also the information given in Sect. 6.1 and 13.
Only the EEx flow sensor may be installed in the hazardous area. The EEx
certified signal converter must be installed outside the hazardous area!

Advertising