3 hardware information and error status, fct. 2.02, 4 hardware test, fct. 2.03 – KROHNE IFC 110 Converter User Manual

Page 69

Advertising
background image

Part C Special applications, functional checks, service, and order numbers Sect. 7.3

05/2003

IFC 110 F

69

7.3

Hardware information and error status, Fct. 2.02


• Before consulting the factory about errors or flow measurement problems, please invoke

Fct. 2.02 HARDW. INFO (hardware information).

• An 8-character and a 10-character status code are stored under this function in each of

3 ”windows”. These 6 status codes allow your compact flowmeter to be subjected to a simple
and rapid diagnosis.

• Switch on the system.

• Press the following keys for a display of the status codes:

Key Display

Description

If ”YES” is selected in Fct. 3.04 ENTRY CODE,

key in 9-stroke CODE 1 now:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

Fct.

1.00 OPERATION

Fct. 2.00

TEST

Fct. 2.01

TEST Q

Fct. 2.02

HARDW. INFO

→ → MODUL ADC

-.- - - - -.- -

1st window

- - - - - - - - -

Example for status code

↵ → MODUL I/O

-.- - - - -.- -

2nd window

3.25105.02 ( 8-character code, 1st line)

- - - - - - - - -

3A47F01DB1 (10-character code, 2nd line)

↵ → MODUL DISP.

-.- - - - -.- -

3rd window

- - - - - - - - -

PLEASE NOTE DOWN ALL 6 STATUS CODES !

Fct. 2.02

HARDW. INFO

Terminate hardware information

(2x) 3x

↵ - - - - - - - -

- - - - - / - - -

Measuring mode


7.4

Hardware test, Fct. 2.03


Please note:
Before beginning the test, deactivate any alarms and controllers as the current output will be
tested with three values 4, 4.7 and 23 mA for a short period.

Key Display

Description

If ”YES” is selected in Fct. 3.04 ENTRY CODE,

key in 9-stroke CODE 1 now:

→ → → ↑ ↑ ↑ ↵ ↵ ↵

Fct.

1.00

OPERATION

Fct. 2.00

TEST

Fct. 2.01

TEST Q

2 x

Fct. 2.03

HARDW. TEST

Hardware test

SURE

NO

SURE

YES

WAIT

Hardware test in progress, duration approx. 60 seconds

- - - - - - - -

1st error

List of errors see Sect. 4.5. Errors are

- - - - - - - -

2nd error

always displayed independent of setting in

- - - - - - - -

3rd error

}

Fct. 1.04. If no error is detected, refer to next line.

Fct. 2.03

HARDW. TEST

Terminate hardware test

(2x) 3x

↵ - - - - - - - -

- - - - - / - - -

Measuring mode



If you need to return your flowmeter to KROHNE, please refer to the last-but-one page of

these instructions.