5 error messages in measuring mode – KROHNE IFC 110 Converter User Manual

Page 38

Advertising
background image

Sect. 4.5

Part B IFC 110 F Signal converter

38

IFC 110 F

05/2003

Fct. Text

Description

and

setting

LIMIT VAL.

Set limit value for noise/interference suppression

(appears only when ”YES” is selected under ”SPEC. FILT.”, see above)

Setting

range: 01-90 PERCENT of full-scale range Q

100%

see Fct. 3.02, subfunction ”FULL SCALE”

Press

key to change to subfunction ”LIMIT CNT.”.

LIMIT CNT.

Totalizer active when exceeding limit value (see ”LIMIT VAL.” above)

(appears only when ”YES” is selected under ”SPEC. FILT.”)

Setting

range: 001-250

Press

key to return to Fct. 3.06 ”APPLICAT.”.

3.07

HARDWARE

Determine HARDWARE functions

TERM.A1

Terminal A1

● PULSOUTP.

● STATUSOUTP.

Select with key

↑.

Press key

to transfer to subfunction ”SELFCHECK”.

SELFCHECK

Carry out self check? See Section 5.18.

● YES

● NO

(testing different parameters)

Press key

to transfer to subfunction “FIELD CURRENT”.

FIELDCUR.

Determine field current

● INTERNAL

● EXTERNAL (only with power driver, see Sect. 8.6)

Press

key to return to Fct. 3.07 ”HARDWARE”.

4.5

Error messages in measuring mode

The following list contains all errors which may occur during flow measurment. Errors are displayed when
”YES” is selected in Fct. 1.04 DISPLAY, subfunction ”DISP. MSG.”.

Error message

Description of error

Elimination of error

LINE INT.

Power failure Note:

Cancel error in RESET/QUIT. menu

no counting during power failure

Reset totalizer if necessary.

OVERFLOW I

Current output overranged.

Check instrument parameters and

or

(flow rate > measuring range)

correct if necessary. After elimination of

OVERFL. I2

the cause, the error message is

cancelled

automatically.

See Sections 6.4 and 6.7.

OVERFLOW P

Pulse output P

Check instrument parameters and

or

or

correct if necessary. After elimination of

OVERFL. P2

Pulse output range P2 exceeded

the cause, the error message is

(flow rate > modulation range)

cancelled automatically.

See Sections 6.4 and 6.7.

I SHORT or

*

Current output I or I2 externally

Check mA loop and increase load using

I2 SHORT

shorted or load < 15

additional resistor if necessary.

I OPEN or

*

mA loop interrupted by current

Check mA loop and reduce load to

I2 OPEN

output I or I2 or load > 500

Ω 500

Ω if necessary.

TOTALIZER

Overflow of internal totalizer

Delete error message in RESET/QUIT

menu, see Sect. 4.6

ADC

Analog/digital converter range

Set Fct. 3.06, subfunction ADC GAIN

exceeded

to ”10”. See Sections 6.4 and 6.7.

If error message does not disappear,

consult

factory.

ADC-PARAM.

Check sum error

Replace ADC printed circuit board

ADC-HARDW.

Hardware error A/D converter

Replace ADC printed circuit board

ADC GAIN

Hardware error A/D converter

Replace ADC printed circuit board

FC-HARDW.

Hardware error on field current PCB

Replace field current PCB

FATAL.ERROR

Fatal error, measurement interrupted

Replace electronic unit or consult factory.

EP PARAM.

Parameters of “Empty Pipe” are wrong

Error shut off automatically EPD function.
Please check calibration values!
Value Empty Pipe – Value Full pipe

≥ 10

Values must be in range of 0 – 150.

*

only for active operation