Sect. 7.1

Part C Special applications, functional checks, service, and order numbers

68

IFC 110 F

05/2003

7 Functional

checks

7.1

Checking the zero with IFC 110 F signal converter, Fct. 3.03


• Set ”zero” flow in the pipeline. Make sure that the measuring tube is completely filled with

liquid.

• Switch on the system and wait 15 minutes.

• Press the following keys for zero measurement:

Key Display

Description

If ”YES” is selected in Fct. 3.04 ENTRY CODE,

key in 9-stroke CODE 1 now:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

Fct.

1.00

OPERATION

2x

Fct. 3.00

INSTALL.

Fct. 3.01

LANGUAGE

2x

Fct. 3.03

ZERO SET

CALIB.

NO

CALIB.

YES

0.00

- - - - - / - - -

Flow rate displayed in set unit, see Fct. 1.04 DISPLAY,

subfunction ”DISP. FLOW”.

Zero is measured, duration approx. 15-90 s. ”WARNING” is

displayed when flow is ”>0”, acknowledge by pressing

↵ key.

STORE

NO

If new value is not to be stored, press

↵ key (3 times)

4 times = return to measuring mode

STORE

YES

Fct. 3.03

ZERO SET

Store new zero value

(2x) 3x

- - - - - - -

- - - - - / - - -

Measuring mode with new zero


7.2

Checking the measuring range Q, Fct. 2.01


• For this test a measuring value can be simulated in the range of -110 to +110 percent of Q

100%

(set full-scale range, see Fct. 1.01 FULL SCALE).

• Switch on the system.

• Press the following keys for checking the measuring range:

Key Display

Description

If ”YES” is selected in Fct. 3.04 ENTRY CODE,

key in 9-stroke CODE 1 now:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

Fct.

1.00

OPERATION

Fct. 2.00

TEST

Fct. 2.01

TEST Q

SURE

NO

SURE

YES

0

PCT.

Current, pulse and status outputs indicate corresponding values.

± 10

PCT.

± 50

PCT.

Select with

↑ key

± 100

PCT.

± 110

PCT.

Fct. 2.01

TEST Q

End of test, actual measured values again available at outputs

(2x) 3x

- - - - - - -

- - - - - / - - -

Measuring mode