6 checking the flow sensor – KROHNE IFC 110 Converter User Manual

Page 76

Advertising
background image

Sect. 7.6

Part C Special applications, functional checks, service, and order numbers

76

IFC 110 F

05/2003

7.6

Checking the flow sensor


Always switch off the power supply before opening the housing.

Required measuring instruments and tools

• Ohmmeter with at least 6 V measuring voltage range

• or AC voltage/resistance bridge

Note:

accurate measurements in the electrode area can only be obtained with
an AC voltage/resistance bridge. The measured resistance also heavily
depends on the electrical conductivity of the process liquid.


Preparations

• Switch off the power supply.

• Remove the cover from the terminal compartment (remove 2 screws).

• Pull out the two plug-in terminals SC (5-pin, signal line) and FP (4-pin, field current supply

line), see illustration in Section 8.1.

• Fill the measuring tube of the flowmeter completely with process liquid.

Please note:

the following measurements must only be carried out for plug-in terminals
which are occupied (used).

Action

Typical

Incorrect result for

Resistance measurements at

result 1–3

= defective flow sensor,

plug-in terminals SC (5-pin, signal line)

return to factory for repair,

and FP (4-pin, field current supply line)

refer to last-but-one page !

1

Measure resistance between wires

30 – 170

If lower,

7 and 8

interwinding

fault.

If

higher,

wire

break.

2

Measure resistance between wires

> 20 M

If lower, interwinding fault

1 and 7 or between wires 1 and 8

to PE or FE.

3

Measure resistance between wires

1 k

Ω – 1 MΩ

If lower, drain measuring tube

1 and 2 and between 1 and 3 (see

”Note” above)

and repeat measurement;

(same measuring conductor always on

Both values should

if still too low, short-circuit in

wire 1 ! )

be approx. equal

electrode wires.

If

higher, break in electrode wires

or

electrodes

contaminated.

If values differ considerably,

break in electrode wires or

electrodes

contaminated.

4

When signal line B (type BTS/

> 20 M

If lower,

bootstrap) is used: line

fault.

measure resistance between

the following lines:

Check connection cables,

1 and 20 / 1 and 30 / 20 and 30

replace signal line if necessary.

2 and 20 / 3 and 30