Part C Special applications, functional checks, service, and order numbers Sect. 7.5

05/2003

IFC 110 F

71

Group D

Display

Cause

Remedial action

D 1

LINE INT.

Power failure

Delete error message in

Note: no counting during

RESET/QUIT. menu,

power failure

reset totalizer if necessary.

D 2

OVERFL. I

Current output range

Check instrument parameters

exceeded

and correct if necessary. Error

message is deleted automati-

cally after cause has been

eliminated.

D 3

OVERFL. P

Pulse output range

Check instrument parameters

exceeded

and correct if necessary. Reset

Note:

totalizer. Error message is dele-

totalizer deviation is

ted automatically after cause

possible

has been eliminated.

D 4

ADW

A/D converter range

Error message is deleted

exceeded

automatically after cause

has been eliminated.

D 5

FATAL.ERROR

Fatal Error, all outputs are

Replace signal converter,

set to ”min” values

see Sect. 8.3 or consult KROHNE

Service, having first noted down

hardware information and error

status, see Sect. 7.3, Fct. 2.02.

D 6

TOTALIZER

Counts lost

Delete error message in

(overflow, data error)

RESET/QUIT. menu.

D 7

I SHORT

Short circuit at current

Check electrical connection

output

acc. to Sect. 2.2 and correct if

necessary.

Load

≥ 15 Ω !

D 8

I OPEN

Open current output

Provide load

≤ 500 Ω !

D 9

ADC PARAM.

Fault detected on the

Check measuring accuracy.

D 10

ADC HARDW.

ADC printed circuit

Replace ADC printed circuit

D 11

ADC GAIN

board (see Sect. 8.4) or con-

sult KROHNE Service, having

first noted down hardware

information and error status,

see Sect.. 7.3, Fct. 2.02

D 12

STARTUP,

Hardware fault,

Replace signal converter

cyclic flashing

watchdog trips

(see Sect. 8.3) or consult

KROHNE Service, having first

noted

down

hardware

information and error status,

see Sect. 7.3, Fct. 2.02.

D 13

BUSY

Displays for flow, totalizers

Change setting in Fct. 1.4

and

messages

disabled

D 14

Unsteady display

Low electrical conductivity,

Increase time constant in

high solids content,

Fct. 1.2, refer to Sect. 6.5 and 6.7.

pulsating

flow

D 15

No display

Power supply OFF

Switch on power supply.

Check power supply fuse F5

Replace if blown,

(F6 for DC versions)

in terminal compartment