Checking the e5810a, Checking front and rear panel displays – Agilent Technologies E5810A User Manual

Page 110

Advertising
background image

110

Chapter 4

Troubleshooting Information
Checking the E5810A

Checking the E5810A

This section gives guidelines to check the E5810A, including:

„

Checking Front and Rear Panel Displays

„

Checking E5810A Configuration Parameters

Checking Front and Rear Panel Displays

As a first step in troubleshooting the system, you should check the front
panel display and LED status. In addition, you can check the rear panel
Ln LED to see if there is communication from your PC to the E5810 via the
LAN.

Checking LEDs

The following table shows suggested steps to troubleshoot the E5810 using
the LED displays. If there is an error message on the front panel display, see

Checking Front Panel Display Error Messages”

.

NOTE

The Fault LED is turned ON briefly during power-on to show it is working
and then turns OFF. If a hardware fault occurs, the Fault LED turns ON
and remains ON for more than 10 seconds. In this case, contact Agilent
Technologies.

LED Status

Possible Cause(s)

Corrective Action(s)

All LEDs OFF

No AC power to the E5810

Check AC power connections.

Only POWER and
FAULT ON

If FAULT is ON for more than 10
seconds, suspect an E5810
hardware failure

Contact Agilent for instructions to
return the E5810.

POWER ON, back
panel Ln OFF

Indicates no communication from
your PC to the E5810 via the LAN.

Check the LAN connection to the
E5810. See

Chapter 1

, “E5810A

Description” for location of the LAN
port and the Ln LED.

Advertising