Typical Telnet

Screen Display

A typical display screen for an E5810 follows. The configuration values you
see for your E5810 will probably be different from those shown.

Welcome to the E5810 LAN/GPIB Gateway Configuration Utility.

Controls GPIB and RS-232 interfaces via the LAN

Commands

?
exit, quit
reboot
status

View Available Commands
Exit WITHOUT Saving Configuration Changes (see reboot)
Save Configuration Changes and Restart E5810
View the LAN/GPIB Gateway Connection Status

Read-only E5810 Parameters

hardware-addr:
serial-num:

0030D305001E
US12341234

# Ethernet (MAC) Address
# Serial Number

Configurable Parameters saved in E5810 non-volatile memory

(Note: Some E5810 current values “in-use” may be different)

dhcp:

ip:
subnet-mask:
gateway:

dns-server
hostname:

OFF
156.140.104.232
255.255.254.0
156.140.104.1
156.140.2.11
my1234567

# Configure LAN for DHCP boot
# Internet Protocol (IP) Address
# Network Subnet Mask
# Network Gateway
# DNS Server
# Internet Hostname

description:

Agilent E5810 (00-30-D3-05-00-1E)
# Device Description (UPnP Friendly Name)

upnp:
lan-timeout:
io-timeout:
gpib-name:
gpib-address:
gpib-unit:

rs232-name:
rs232-baud:
rs232-bits:
rs232-stopbits:
rs232-parity:
rs232-flow:
rs232-srq:

ON
600
120
gpib0
21
7

COM1
9600
8
1
NONE
NONE
RI

# Configured as UPnP device
# LAN Timeout (Keepalive) in sec
# I/O Timeout in seconds
# GPIB SICL Interface Name
# GPIB System Controller Address
# GPIB Logical Unit (LU) Number

# RS-232 SICL Interface Name
# RS-232 Baud Rate
# RS-232 Bits
# RS-232 Stop Bits
# RS-232 Parity
# RS-232 Flow Control
# RS-232 SRQ