Chapter 4

Troubleshooting Information
Checking the Network

SICL Timeout

Set to 0 (infinite)

The E5810 uses the LAN maximum timeout configured in Connection
Expert’s LAN interface for its default or infinite timeout. If a SICL
itimeout(id,0)

command is used (where a 0 timeout value is an infinite

timeout) the E5810 will use the

I/O Timeout (sec):

value set on the

Current

Configuration of E5810 LAN/GPIB Gateway

page. The default timeout value is

120 sec. Setting the

I/O Timeout (sec):

value to 0 enables infinite timeouts

from SICL applications.

Setting long timeouts with the E5810 is discouraged. However, if a long
timeout period is required for your application, you can set the timeout with
these commands:

VISA: viSetAttribute(vi, attr, attrState)

where: attr is VI_ATTR_TMO_VALUE

attrState is a large value

SICL: itimeout(id,tval), where tval is a large

number

NOTE

It is highly recommended you

not

set an infinite timeout which can cause

the E5810 to become unresponsive due to a pending operation that will
never timeout.