Agilent Technologies E5810A User Manual

Page 125

Advertising
background image

Chapter 4

125

Troubleshooting Information

Checking PC Client Connections

viopen

fails or I/O

operation fails: Out
of resources

The E5810 was unable to open another session or perform the
operation because it is out of resources. You may need to have a client
terminate its connection or you may need to reboot the E5810 to close
all client connections to reclaim resources. Rebooting will disconnect all
other users accessing the E5810.

I/O operation fails:
No connection

The connection to the E5810 failed because:

„

The network connection is down or the E5810 is down and a

network timeout occurred.

„

The E5810 was powered off and then on or an E5810 reboot was

executed, either of which terminates any open client connections.

I/O operation fails:
Timeout

There are several possible causes:

„

The client may be attempting a transaction to a non-existent or

powered-off device. Check the address used, as well as the
status of the device.

„

The network connection or the E5810 has gone down since the

previous I/O operation.

„

The I/O timeout specified is not long enough for the transaction to

complete. You may need to verify that the E5810 I/O timeout
configuration value is long enough if the client specified a timeout
value of infinity.

„

The E5810 is busy performing an I/O operation for another client

or another client has the device/interface locked. See “E5810
appears to be hung” for more information.

Message

Description/Actions