Checking the E5810A

Checking E5810A Parameters

If the front panel display and LEDs indicate no error conditions, check the
E5810 configuration parameters.

Checking E5810

Current Parameters

If the E5810 is connected to an Enterprise (corporate) network, you can
check the E5810 current settings displayed on the

Configuring your E5810

LAN/GPIB Gateway

page against the Enterprise Network Information

provided to you by your System Administrator. If the settings do not match,
re-configure the E5810 from the

Configuring your E5810 LAN/GPIB Gateway

page.

Resetting E5810

Defaults

As required, you can use the

Preset

button on the E5810 front panel to reset

the E5810 to its default settings. There are two modes of operation. See

Chapter 1

, “E5810A Description” for the location of the

Preset

button.

„

Momentarily Depress the Preset Button.

If you momentarily

depress the

Preset

button, ONLY the password is temporarily

set to the default value of E5810. All other E5810 configuration
parameters remain the same. The password remains until the
next time you reboot the E5810.

„

Depress the Preset Button for at Least 10 Seconds.

If you

depress the

Preset

button for at least 10 seconds, ALL E5810

configuration parameters are set to their defaults and the
E5810 reboots.

NOTE

If the E5810 is connected to an Enterprise (corporate) Network, be sure
to contact your IT Department before making any configuration changes
to the E5810. If E5810 configuration checks are required, see

Chapter 3

, “Using E5810A Web Access” for information.

NOTE

You should terminate all applications that use the E5810 before taking
this action.