To create a custom multitone waveform – Agilent Technologies E8247C PSG CW User Manual

Page 181

Advertising
background image

Chapter 8

171

Multitone Waveform Generator

Creating, Viewing, and Optimizing Multitone Waveforms

Creating, Viewing, and Optimizing Multitone Waveforms

This section describes how to set up, generate, and optimize a multitone waveform while viewing it with a
spectrum analyzer. Although you can view a generated multitone signal using any spectrum analyzer that
has sufficient frequency range, an Agilent Technologies PSA high-performance spectrum analyzer was used
for this demonstration. Before generating your signal, connect the spectrum analyzer to the signal generator
as shown in

Figure 8-1

.

Figure 8-1

Spectrum Analyzer Setup

To Create a Custom Multitone Waveform

Using the

Multitone Setup

table editor, you can define, modify and store user-defined multitone

waveforms. Multitone waveforms are generated by the dual arbitrary waveform generator.

1. Preset the signal generator.

2. Set the signal generator RF output frequency to 20 GHz.

3. Set the signal generator RF output amplitude to 0 dBm.

4. Press

Mode

>

Multitone

>

Initialize Table

>

Number of Tones

>

9

>

Enter

.

5. Press

Freq Spacing

>

1

>

MHz

.

6. Press

Initialize Phase Fixed Random

to Random.

7. Press

Done

.

8. Press

Multitone Off On

to On.

9. Turn on the RF output.

Advertising
This manual is related to the following products: