To determine peak to average characteristics – Agilent Technologies E8247C PSG CW User Manual

Page 187

Advertising
background image

Chapter 8

177

Multitone Waveform Generator

Creating, Viewing, and Optimizing Multitone Waveforms

To Determine Peak to Average Characteristics

This procedure describes how to set the phases of the tones in a multitone waveform and determine the peak
to average characteristics by plotting the complementary cumulative distribution function (CCDF).

1. Press

Mode > Multitone > Initialize Table

>

Number of Tones

>

64

>

Enter

.

2. Press

Freq Spacing

>

20

>

kHz

.

3. Press

Initialize Phase Fixed Random

to Fixed.

4. Press

Done

.

5. Press

Apply Multitone

.

6. Press

More (1 of 2) > Waveform Statistics > Plot CCDF

.

You should now see a display that is similar to the one shown in

Figure 8-7

. The CCDF plot displays the

peak to average characteristics of the waveform with all phases set to zero.

Figure 8-7

CCDF Plot with Fixed Phase Set

7. Press

Mode Setup > Initialize Table

.

8. Press

Initialize Phase Fixed Random

to Random.

9. Press

Random Seed Fixed Random

to Random.

Peak