Agilent Technologies E8247C PSG CW User Manual

Page 48

Advertising
background image

38

Chapter 2

Basic Operation
Configuring the RF Output

Configuring a Frequency Sweep

1. Set up the equipment as shown in

Figure 2-2

.

NOTE

The PSG signal generator is not compatible with the GPIB system interface of an 8757A,

8757C, or 8757E. For these older scalar network analyzers, do not connect the GPIB
cable in

Figure 2-2

. This method provides only a subset of 8757D functionality. See the

PSG Data Sheet for details. Use the 8757A/C/E documentation instead of this
procedure.

Figure 2-2

Equipment Setup

2. Turn on both the 8757D and the PSG.

3. On the 8757D, press

SYSTEM

>

MORE

>

SWEEP MODE

and verify that the

SYSINTF

softkey is set to ON.

This ensures that the system interface mode is activated on the 8757D. The system interface mode
enables the instruments to work as a system.

4. Press

Utility

>

GPIB/RS-232 LAN

to view the PSG’s GPIB address under the

GPIB Address

softkey. If you

want to change it, press

GPIB Address

and change the value.

5. On the 8757D, press

LOCAL

>

SWEEPER

and check the GPIB address. If it does not match that of the PSG,

change the value.