3 optimizing performance, Selecting alc bandwidth, To select an alc bandwidth – Agilent Technologies E8247C PSG CW User Manual

Page 69: Optimizing performance

Advertising
background image

59

3

Optimizing Performance

In the following sections, this chapter describes procedures that improve the performance of the Agilent PSG
signal generator.

Selecting ALC Bandwidth

(below)

“Using External Leveling” on page 60

“Creating and Applying User Flatness Correction” on page 64

“Adjusting Reference Oscillator Bandwidth (Option UNR)” on page 76

Selecting ALC Bandwidth

For internal leveling, the signal generator uses automatic leveling control (ALC) circuitry prior to the RF
output. ALC bandwidth has five selections: automatic, 100 Hz, 1 kHz, 10 kHz, and 100 kHz. In automatic
mode (the preset selection), the signal generator automatically adjusts the ALC bandwidth among three of
the four possible settings, depending on the active functions (see

Figure 3-1

).

Figure 3-1

Decision Tree for Automatic ALC Bandwidth Selection

To Select an ALC Bandwidth

Press

Amplitude

>

ALC BW

>

100 Hz

,

1 kHz

,

10 kHz

, or

100 kHz

.

This overrides the automatic ALC bandwidth selection with your specific selection.

RF OUTPUT

< 2 MHz

AM OFF

PULSE OFF

AM OFF

PULSE ON

AM ON

PULSE ON

AM ON

PULSE OFF

ALC BW

100 Hz

ALC BW

1 kHz

ALC BW

10 kHz

ALC BW
100 kHz

Yes

No

No

No

No

Yes

Yes

Yes

Yes

Advertising