Vaisala DM500 User Manual

Page 102

Advertising
background image

User's Guide ______________________________________________________________________

100 _________________________________________________________________ M210205EN-D

XFLOW Setting the Flow Value Temporarily

Syntax:

XFLOW x.xxx<cr>

where
x.xxx =

sample air flow value

The function of XFLOW is the same as with the command FLOW
except that with the command XFLOW the setting is valid only until
power is turned off or the command XFLOW is given without
parameters. After this the flow rate stored with the command FLOW
is valid again.

PRES Setting the Pressure

Syntax:

PRES<cr>

Sets the pressure if there is no pressure transducer in the system. The
pressure value is needed in volume flow rate calculations in the T

d

measuring algorithm as well as in T

d

→ ppm unit conversion.

Example:

>pres ?
Pressure : 1.0132 bar
>pres 1.0
Pressure : 1 bar
>

NOTE

If the pressure setting is frequently adjusted, e.g. by using an external
barometer as a pressure input source, the command XPRES is
recommended.

XPRES Setting the Pressure Temporarily

Syntax:

XPRES xx.xx<cr>

where
xx.xx =

sample air pressure value