Response speed, Pump on/off – Vaisala DM500 User Manual

Page 64

Advertising
background image

User's Guide ______________________________________________________________________

62 __________________________________________________________________ M210205EN-D

Response Speed

You can set the response speed to stable, normal (default) or fast. If
fast transients in the dewpoint prevent the system from reaching
balance, the signal can be stabilized by selecting the Stable response
speed. Fastest response time is achieved by setting the speed to Fast
but this may increase the noise in the dewpoint measurement. As a
rule of thumb, always use the fastest setting that gives stable readings!

1.

Open

MENU

: press .

2.

Select ►

Measuring

, press .

3.

Select ►

Measurement settings

, press .

4.

Select

Response

, press

SET

.

5.

Select one of the following options:

Stable

: Press

SELECT

to have the slowest response speed.

Normal

: Press

SELECT

to have normal response speed.

Fast:

Press

SELECT

to have fast response speed.

6.

Press

EXIT

to return to the main view.

Pump ON/OFF

1.

Open

MENU

: press .

2.

Select ►

Measuring

, press .

3.

Select ►

Measurement settings

, press .

4.

Select

Pump

, press

ON/OFF

.

5.

Press

EXIT

to return to the main view.