General – Vaisala DM500 User Manual

Page 123

Advertising
background image

Chapter 19 ____________________________________________________________ Specifications

VAISALA ______________________________________________________________________ 121

External Temperature (Optional)

Sensor

Pt100RTD DIN IEC 751 class ¼B

Measurement range

-40...+80 ºC

Typical accuracy at +20 ºC

±0.1 ºC

Typical temperature dependence
of electronics

0.001 ºC

Cable length

2 m/5 m/10 m

Integrated Pressure (Optional)

for ppmv calculations
Measurement range

0...2 bar or 0... 20 bar

Accuracy

±1.0 % of full scale

Integrated Flow (Optional)

Measurement range

0...2 l/min

Accuracy

±10 % of reading

Integrated flow adjustment

General

DMP501 Dewpoint Sensing Unit

Operating voltage

12 VDC, 25 A

(when used stand alone)
Operating temperature

-10...+60 ºC

Pressure of sample gas

0...20 bar

Sample gas flow rate

1.0 slpm

Cooling water temperature

0...15 ºC

Operating gases

Air, N2, Ar, SF6;
nontoxic,
noncorrosive,
nonflammable gases
only.

Options may have extra restrictions.

Electronics housing material

aluminium, stainless
steel

Advertising