Chapter 12 ___________________________________________________ Self Diagnostics Settings

VAISALA _______________________________________________________________________ 67

the sensor is desired, you can set the automatic checking interval
(1...99 h) as follows:

1.

Open

MENU

: press .

2.

Select ►

Measuring

, press .

3.

Select ►

Self diagnostics

, press .

4.

Select

Sensor circuit chk.

5.

To turn on the sensor checking circuit and set the checking
interval, press

SET

.

6.

Set the interval (1...99 h) by using the numerical buttons.

7.

Press

OK

to confirm the setting. Now the automatic sensor

circuit check will be done in determined intervals.

8.

If needed to turn off the checking operation, press

OFF

.

9.

Press

EXIT

to return to the main view.

Cooling Capacity Test

In the cooling capacity test the sensor is cooled to the lowest possible
temperature. The measurement mode must be stopped during the
cooling capacity test.

1.

Open

MENU

: press .

2.

Select ►

Measuring

, press .

3.

Select ►

Self diagnostics

, press .

4.

Select

Cooling capacity test

5.

Press

START

to start the test. The test may take up to 20 min.

During that time the measuring is stopped and you can follow
the actual temperature of the sensor from the display: