Calibrating DM500 Equipped with a
Pump

If you have the DM500 with a pump, ensure that the sample flow rate
from the humidity generator is sufficient. See the following example:

Humidity
generator

Gas flow
> 3 slpm

1 splm

1 splm

DM500

Reference
instrument

Open outlet
(with diffusion coil)

Figure 22

Gas Flow Rate from the Humidity Generator

Making an Adjustment

DMP501 needs 4...10 dewpoint reference points within dewpoint
temperature range -80 ºC...+60 ºC for the adjustment. These dewpoint
reference points shall be traceable to appropriate standards. The
dewpoint temperature references shall cover the measurement area in
which the instrument is used.

The adjustment can be done also by using the serial commands.

If you have adjusted the device earlier, you shall clear the previous
adjustments before giving new adjustment data.

1.

Open

MENU

: press .

2.

Select ►

Settings

, press .

3.

Select ►

Adjustments

, press .

4.

Select ►

Dewpoint adjustments

, press .

5.

Press CLEAR and YES , OK to clear the old adjustment
data

Connect the reference gas flow to the gas inlet and measure the
readings in 4...10 points. Write down the readings in the different
reference points. Feed the reference points and corresponding readings
to the device to make the adjustment.