Chapter 10 __________________________________________________________ Display Settings

VAISALA _______________________________________________________________________ 57

1.

Open

MENU

: press .

2.

Select ►

Display

, press .

3.

Select

Quantity

, press

SET

, select the quantity for the alarm and

press

SELECT

.

4.

Select

Alert above

, press

SET

. (In case you want to delete the

setpoint, press

DELETE

and go to item 6).

5.

Press

CLEAR

and give the value by using numerical buttons.

Press

OK

.

6.

Select

Alert below

, press

SET

. (In case you want to delete the

setpoint, press

DELETE

and go to item 8).

7.

Press

CLEAR

and give the value by using numerical buttons.

Press

OK

.

8.

Select

ENABLED WHEN

, press

SELECT

. Now you can select the

measurement status that activates the alarm.

STOP

: alarm is enabled when dewpoint measuring status is STOP.

ERR

: alarm is enabled when dewpoint measuring status is ERR.

OK

: alarm is enabled when dewpoint measuring status is OK

WAIT

: alarm is enabled when dewpoint measuring status is WAIT

All the modes selected: alarming is not dependent on any
measurement status (only on numeric set points).

None of the modes selected (never): alarming is disabled.

You can select any combination of the measurement status to control
the alarm.

9.

Press

OK

and

EXIT

to return to the main view.

Alarming

The display is beeping and blinking and the note sign message is
shown in the display. Mute the alarming by pressing

MUTE

and

OK

.

If the alarming condition is still valid a small note sign remains in the
header of the display. The note sign disappears when the alarming
condition is not valid anymore.