Chapter 16 _____________________________________________________ Serial Line Commands

VAISALA ______________________________________________________________________ 107

HELP Outputting the Commands

Syntax:

HELP<cr>

Example:

>help
Available commands:
ADDR CLOSE COOLER FLOW ICE INIE INTV
OPEN PRES PUMP R SALT SEND SMODE
SNUM START STOP TIME TP UNIT VERS
ECHO ALARM ASEL
SERI FORM ERRS ? DIR DEL PLAY
HELP

INIE Reverting the Factory Settings

Syntax:

INIE<cr>

Does not affect calibration, date, time or data file memory.

Example:

>inie
All user settings will be deleted!
Are you sure : (Y/N) ? Y
OK

SMODE Selecting the Operation Mode

Syntax:

SMODE x<cr>

where
x

=

STOP, RUN, or POLL

The SMODE command is used to set or inspect the default operation
mode of the EIA-232 (and optional RS485) interfaces.

STOP mode: After power-up the transmitter outputs its type and
software version and then waits for further commands.

RUN mode: Continuous output starts automatically from power-up.
Output can be defined with the command FORM. Only command S
can be used.

POLL mode: Allows the communication with multiple transmitters
or other digital instruments connected to one serial bus. Only the