Chapter 14 _________________________________________________________ Output Interfaces

VAISALA _______________________________________________________________________ 77

WAIT: output is enabled when dewpoint measuring status is
WAIT

All the modes selected: output activity is not dependent on any
measurement status.

None of the modes selected (never): output is not enabled
(invalid output).

You can select any combination to control the output.

12.

Select

Invalid output

to set the output reading when the output is

not enabled. Press

SET

(setting can be done also by using serial

line, see command AERR).

13.

Press

CLEAR

and give the invalid value by using numerical

buttons. The value of the invalid output shall be 4...20 mA.
Press

OK

.

14.

Press

EXIT

to return to the main view.

15.

Set the other outputs similarly.

Analog outputs can be configured also by using the serial line (see
command ASEL).

Relay Outputs (Optional)

The DMI500 user interface can be equipped with two configurable
single pole double throw relays. The maximum voltage connected to
relay terminals is 42 V DC/60 V Peak. Select the quantity, setpoints,
and hysteresis and set the criteria for enabling the function.

1.

Open

MENU

: press .

2.

Select ►

Interfaces

, press .

3.

Select ►

Relay outputs

, press .

4.

Select ►

Relay 1

, press .

5.

Select

Quantity

, press

SET

, select the quantity for the relay

output and press

SELECT

.

6.

Select

Active above: xx

, press

SET

. Set the value above which you

want the relay to be activated. Use the numeric buttons. Press

OK

.