Figure 9, Figure 10 – Vaisala DM500 User Manual

Page 31

Advertising
background image

Chapter 4 _________________________________ General Information About Sampling and Results

VAISALA _______________________________________________________________________ 29

The Effect of the Sample Gas Flow Rate on
Cooling Capacity and Result Deviation

The cooling effect of the gas flow rate is
shown in

Figure 9

(on the right). Low flow

rate increases cooling capacity but also
increases response time.

-79

-78

-77

-76

-75

-74

-73

-72

0

0.5

1

1.5

2

2.5

Flow l/min

Tsensor /C

Figure 9

Effect of Sample Flow Rate on Lowest
Achievable Sensor Temperature

The sample flow rate has an effect on the
deviation of the measurement results as
shown in

Figure 10

(on the right).

Sample flow rate range: 0.5...1.5 slpm.

Recommended sample flow rate: 1.0 slpm.

-0.2

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

0.5

0.75

1

1.25

1.5

flow/slpm

devi

ati

on/

°C

Figure 10

Effect of Sample Flow Rate on
Dewpoint Measurement at -60 ºC