Appendix A __________________________________________________________ Humidity Theory

VAISALA ______________________________________________________________________ 127

APPENDIX A

HUMIDITY THEORY

This Appendix provides you with a basic explanation of the terms
related to humidity theory that are used in many parts of this manual.

Measured and Calculated Results of DM500

DM500 measures the dewpoint temperature of a gas by condensing
water on a surface and using SAW technology to detect this
condensation. The temperature at which condensation occurs is
accurately measured by a four wire platinum resistance temperature
sensor and is reported as dewpoint/frostpoint temperature. Other
humidity parameters are calculated by using the measured
dewpoint/frostpoint temperature data and either measured or set
values for pressure and temperature.

Water Vapor in Air

Water vapor is water in the gaseous phase. It is an abundant
component of the earth’s atmosphere, and it is also common in many
industrial processes. At atmospheric pressure, water behaves like an
ideal gas.

Dalton’s law states that in a gas mixture, such as air, the total pressure
of the gas is the sum of the partial pressures of each of the component
gases

others

water

oxygen

nitrogen

total

P

P

P

P

P

+

+

+

=

(1)

Thus, atmospheric pressure is actually the sum of the partial pressures
of nitrogen, oxygen, water vapor, carbon dioxide, and small amounts
of other gases.