Setting commands – Vaisala DM500 User Manual

Page 97

Advertising
background image

Chapter 16 _____________________________________________________ Serial Line Commands

VAISALA _______________________________________________________________________ 95

Setting Commands

SERI Setting the Serial Communication
Parameters

Syntax:

SERI [baud] [parity] [bits] [stops]

[duplex]<cr>

where
[baud]

=

300/600/1200/2400/4800/9600/19200/38400/57600/1
15200

[parity]

=

N (none)/E (even)/O (odd)

[bits]

=

7 (7 bits)/ 8 (8 bits)

[stops]

=

1 (1 stop bit)/2 (2 stop bits)

[duplex] =

H(half-duplex)/F (full-duplex)

The current settings are output if you give the command without the
parameter fields. Factory default: 9600/8/none/1.

The fields can be given in any order. The new settings are valid
immediately after the change. If you choose a half-duplex mode, the
echo turns automatically into OFF-position.

>seri 19200 n 8 1 f
19200 N 8 1 FDX
>

UNIT Setting the Units

Syntax:

UNIT [x]<cr>

where
x

=

unit C or F (dewpoint, temperature, wet-bulb temperature),
g/m3 or gr/ft3 (absolute humidity), g/kg or gr/lb (mixing
ratio), kJ/kg or btu/lb (enthalpy), hPa or torr or psi or bar
(pressure), slpm or scfh (sample flow).

x

=

m (metric)or n (non-metric)

The unit is set by typing the desired unit(s) after the command UNIT.
All the units change form metric to non-metric (or other way round)
by typing UNIT m or UNIT n. The display units change
simultaneously.