12.1 Preparation (Setting Parameters)

Step 1

Turn off the display.
Press the

ON/OFF

switch while the

ZERO

switch

is pressed and held. The function table is displayed.

Step 2

Press the

ENTER

switch to display an item of

the comparator function (

f6

).

Step 3

Select a parameter of the full/dribble batch
function (

f6 9

) with the

^

switch.

Step 4

Press the

ENTER

switch to store the new

parameter and display an item of the full/dribble
batch sub-function (

f10

) .

Step 5

Select a parameter of the full/dribble batch sub-
function (

f10 0

~

f10 3

) with the

^

switch.

Step 6

Press the

ENTER

switch to store the new

parameter

Step 7

Press the

F

switch and the

ENTER

switch to

exit from the function table. Then the scale dis-
plays the weighing mode.