16 RS-232C Serial Data Interface

RS-232C interface has the following two modes.
Stream mode

Outputs data countinuously and can be used for printing data.

Command mode Controls the scale using commands from a computer.

It is necessary to set the parameters of the “Baud rate (

f4

)”,

”Output mode (

f5

)”, “ Format (

f15

)” in the function table in advance.

There are option cables as follows:
AX-KO557A RS-232C cable, D-sub 25 pin, 2m
AX-KO1786-200 RS-232C cable, D-sub 9 pin, 2m

Transmission system

EIA RS-232C

Transmission

Asynchronous, bi-directional, half-duplex

Data format

Baud rate: 2400, 4800, 9600 bps
Data:

7 bits

Parity:

1 bit, Even

Start bit

1 bit

Stop bit

1 bit

Code

ASCII

Terminator

CR LF (CR: 0D[H], LF: 0A[H])

Pin Connections

Pin No. Signal name

Direction Description

1

-

2

RXD

Input

Receive data

3

TXD

Output

Transmit data

4

-

5

SG

-

Signal ground

6

-

7

DSR

Output

Data set ready

Circuit